Type R / Type S Precision Thermocouple

The precision thermocouples are used as transfer standards, for high-precision temperature measurements or as calibration standards. The type R / type S thermocouples are precision noble metal thermocouples which can be assembled with a reference junction or with mini plug. They are characterized by the highest available purity of the thermocouple wires.

Highlights

  • Temperature range: 0°C to 1200 °C
  • highest available purity of the thermocouple wires
    The thermocouple wires are specifically thermally treated.
  • small measurement uncertainties with DAkkS calibration certificate
    The thermometer can be calibrated at fixed temperature points with smaller measurement uncertainties.

Measurement Uncertainties

Technical data

Model No. KK-170
Temperature range 0 °C to 1200 °C
Connection   temperature resistant – Thermoelectric cable with mini plug or
very low EMF copper
Thermowire
  • Type R / Type S
  • maximum purity available
  • thermally pre-treated and annealed
Sheath
  • made of gas-tight high-purity ceramic
Handle maximum continuous temperature 100 °C
Function  DIN EN 60584 (individual parameters are calculated during calibration)
Uncertainties of measurement (with DAkkS calibration)
Fixed point calibration
0,01 °C
419,527 °C
660,323 °C
961,78 °C
 

0,2 K
0,5 K
0,5 K
0,6 K

Comparison calibration
0 °C to 962 °C
>962 °C to 1200 °C
 

0,8 K
1,5 K

Dimensions
Protection Sheath Ø 7 mm x 500 mm
Handle Ø 20 mm x 95 mm
Length of thermocouple wire 2 m
cold junction  Ø 6 mm x 280 mm (stainless steel)
Length of connection cable two meter connection cable made of very low EMF copper
Carrying case included (Exclusive wooden case optional)

 

Questions about the thermometer?