Type B Precision-Thermocouple
The precision thermocouples are used as transfer standards, for high-precision temperature measurements or as calibration standards at high temperatures up to 1500°C.
Highlights
- Temperature range: 0°C to 1500°C
- Highest available purity of the thermocouple wires
- The thermocouple wires are specifically thermally treated
- small measurement uncertainties with DAkkS calibration certificate
The thermometer can be calibrated at fixed temperature points with smaller measurement uncertainties
Measurement Uncertainties
Technical Data
Model No. | KK-185 | |||
Connection | thermocouple cable with mini-plug or low EMF copper wire | |||
Type | Type B | |||
Temperature range | 0°C to 1500°C | |||
Thermocouple wire |
|
|||
Measuring point |
|
|||
Handle | maximum continuous temperature 100 °C | |||
Characteristic | DIN EN 60584 (individual parameters are calculated during calibration) |
|||
Uncertainties of Measurement (for a DAkkS calibration) | ||||
Fixed point calibration | ||||
0,01 °C 419,527 °C 660,323 °C 961,78 °C |
0,2 K 0,5 K 0,5 K 0,6 K |
|||
Comparison calibration | ||||
0 °C bis 962 °C > 962 °C bis 1200 °C |
0,8 K 1,5 K |
|||
Dimensions | ||||
Protection tube Measuring point | Ø 7 mm (Length depending on configuration) | |||
Handle | Ø 20 mm x 95 mm | |||
Length thermocouple wire | 2 m | |||
Cold junction | Ø 6 mm x 280 mm (stainless steel) | |||
Length of connecting cable | none or low EMF copper (depending on configuration) | |||
Carrying Case | Case (high quality wooden case optional) |